Objectives: The aim of this study is to investigate the interference of carbamylated haemoglobin (cHb) to haemoglobin A1c (HbA₁c) measurements in patients with chronic kidney disease (CKD) by three ion-exchange chromatography methods.

Design And Methods: We included three ion-exchange chromatography methods (Bio-Rad Variant II, G7 and HA-8160) and one affinity chromatography method (Primus Ultra2) to test HbA₁c. In vitro carbamylation of haemoglobin was also performed to test interference of carbamylation.