The objective of this study was to undertake an in vivo cross calibration of body composition, whole body bone mineral content (BMC) and bone mineral density (BMD) between a Hologic QDR2000 and a GE Healthcare Lunar Prodigy. Twenty-one subjects attending for routine bone densitometry were recruited to the study (19 female and 2 male, aged 30-79 yr). Phantom cross calibrations were carried out using the Bio-Imaging Variable Composition Phantom (VCP) for percentage fat (%fat) and the Bona Fide Phantom (BFP) for BMD.
View Article and Find Full Text PDFIdentification of vertebral fracture has become increasingly important in the diagnosis and management of osteoporosis. This study compares the morphometric techniques on a fan beam dual-energy X-ray absorptiometry (DXA) GE-Lunar Expert system (Expert) using a supine lateral position and a narrow fan beam GE-Lunar Prodigy system (Prodigy; GE Lunar, Madison, WI) that requires lateral decubitus positioning. Patient acceptability, image quality, observer, and equipment variability were determined.
View Article and Find Full Text PDFThe purpose of this study was to develop cross calibrations when replacing three dual-energy X-ray absorptiometers (GE Lunar DPXL [Madison, WI], DPXL, GE Lunar Expert [Madison, WI], Expert, Hologic QDR2000 [Waltham, MA]) with two new GE Lunar Prodigy instruments. Subjects previously scanned on the Expert or QDR2000 were transferred to Prodigy 1 and those previously scanned on the DPXL to Prodigy 2. A cohort of subjects was recalled for each old instrument, and approximately 20 subjects had lumbar spine and hip scans on each old instrument and the appropriate new instrument.