Ann Biol Clin (Paris)
December 1992
During a multicenter evaluation, 16 methods for creatinine measurement have been tested according to the guidelines of the Société française de biologie clinique (SFBC) protocol. Kinetic Jaffé methods, widely used in France, performed on different analytical systems (Astra Beckman, IL 508, RA 1000 Technicon, Hitachi 704, 705, 717 Boehringer, Fara Roche, Progress Kone, Kem-O-Mat Coulter, Perspective France Monitor) have been compared to a continuous flow method with aqueous standards, to enzymatic methods using creatinine amidohydrolase with a colorimetric measurement (Boehringer and Ektachem Kodak) and to an HPLC method. Reproducibility, estimated with four different control sera, proved to be unsatisfactory in some cases as compared to current criteria for imprecision (less than +/- 10 mumol/l for intralaboratory and less than +/- 20 mumol/l for interlaboratory imprecision).
View Article and Find Full Text PDF