Low-energy X-ray sources that operate in the kilovoltage energy range have been shown to induce more cellular damage when compared to their megavoltage counterparts. However, low-energy X-ray sources are more susceptible to the effects of filtration on the beam spectrum. This work sought to characterize the biological effects of the Xoft Axxent® source, a low-energy therapeutic X-ray source, both with and without the titanium vaginal applicator in place.
View Article and Find Full Text PDFPurpose: The purpose of this work was to provide guidance for the lack of an air-kerma rate standard for the S7600 Xoft Axxent® source by providing a correction factor to apply to the National Institute of Standards and Technology (NIST) traceable S7500 well chamber (WC) calibration coefficient before the development of an S7600 standard at NIST.
Methods And Materials: The Attix free air chamber (FAC) at the University of Wisconsin Medical Radiation Research Center was used to measure the air-kerma rate at 50 cm for six S7500 and six S7600 sources. These same sources were then measured using five standard imaging HDR1000+ WCs.
Purpose: The purpose of this work is to provide measured data for the modified TG43 parameters [DeWerd et al.] for the newest, Galden-cooled S7600 Xoft Axxent source model.
Methods: The measurement of radial dose distributions at distances of 1 cm to 4 cm from the source was performed using TLD100 microcubes, EBT3 film, and an Exradin A26 microionization chamber.
Purpose: The purpose of this work was to evaluate differences in air-kerma rate of the older, S7500 water-cooled Xoft Axxent source and newer, S7600 Galden-cooled source.
Methods And Materials: The Attix Free Air Chamber (FAC) at the UWMRRC was used to measure the air-kerma rate at 50 cm for six S7600 Xoft Axxent sources. The average measured air-kerma of the S7600 sources was compared with the measured average air-kerma rate from five S7500 sources.
Purpose: It is important to check stability of ionization chambers in between regular calibration cycles. Stability checks can include individual Co irradiations, use of a beta-emitting check source, or redundant measurements in megavoltage photon beams. While Co irradiators are considered stable, they are rarely found in the clinical setting.
View Article and Find Full Text PDF