Purpose: To investigate the frequency, magnitude and possible causes of frame-shifts that may occur between treatment planning and treatment delivery when performing Gamma Knife radiosurgery with rigid frame-based immobilization.
Methods: Differences between computed tomography (CT) framed fiducial stereotactic coordinate reference and cone beam computed tomography stereotactic coordinates after image registration were recorded for 49 frame-based GK radiosurgery cases performed using the Gamma Knife Icon. Parameters recorded include rotational shifts, translational shifts, and the GK-computed Maximum Shot Displacement (MSD) between the two stereotactic coordinate spaces.
Patient- and provider-reported outcomes are recognized as important in evaluating quality of care, guiding health care policy, comparative effectiveness research, and decision-making in radiation oncology. Combining patient and provider outcome data with a detailed description of disease and therapy is the basis for these analyses. We report on the combination of technical solutions and clinical process changes at our institution that were used in the collection and dissemination of this data.
View Article and Find Full Text PDF