Publications by authors named "Asa Carlsson-Tedgren"

Article Synopsis
  • - The study aimed to improve cervical brachytherapy treatment planning by using a method called inverse treatment planning with pseudo-structures, hypothesizing it could achieve better dose distributions while meeting clinical standards.
  • - Data from 16 previous patients was analyzed, comparing manual planning, straightforward inverse methods, and the pseudo-structures approach, with results indicating varying doses for the clinical tumor volume across the different methods.
  • - The findings showed that using pseudo-structures effectively shaped the dwell time distribution and complied with clinical requirements, confirming its efficacy across two different planning systems.
View Article and Find Full Text PDF

Background: Model-based dose calculation algorithms (MBDCA), such as the Advanced Collapsed cone Engine (ACE) in Oncentra Brachy® can be used to overcome the limitations of the TG-43 formalism. ACE is a point kernel superposition algorithm that calculates the total dose separated into primary, first-scatter, and multiple-scatter dose. Albeit ACE yields accurate results under most circumstances, several studies have reported underestimations of the dose to cortical bone.

View Article and Find Full Text PDF

Aim: Ru eye plaque brachytherapy (BT, interventional radiotherapy) is an eye-preserving treatment for uveal melanoma performed in about 100 clinics worldwide. Despite this relatively low number, there is a considerable variation in clinical practice. In 2022, the BRAPHYQS and Head & Neck and Skin GEC-ESTRO working groups conducted a survey to map the current clinical practice.

View Article and Find Full Text PDF

Background: Inorganic scintillation detectors (ISDs) are promising for in vivo dosimetry in brachytherapy (BT). ISDs have fast response, providing time resolved dose rate information, and high sensitivity, attributed to high atomic numbers. However, the conversion of the detector signal to absorbed dose-to-water is highly dependent on the energy spectrum of the incident radiation.

View Article and Find Full Text PDF

The introduction of model-based dose calculation algorithms (MBDCAs) in brachytherapy provides an opportunity for a more accurate dose calculation and opens the possibility for novel, innovative treatment modalities. The joint AAPM, ESTRO, and ABG Task Group 186 (TG-186) report provided guidance to early adopters. However, the commissioning aspect of these algorithms was described only in general terms with no quantitative goals.

View Article and Find Full Text PDF

Purpose: To develop a prognostic score that correlates to a low, medium, and high incidence of treatment failure after plaque brachytherapy of uveal melanoma (UM).

Methods And Materials: All patients who have received plaque brachytherapy for posterior UM at St. Erik Eye Hospital in Stockholm, Sweden from 1995 through 2019 were included (n = 1636).

View Article and Find Full Text PDF

Purpose: The aim was to evaluate a postprocessing optimization algorithm's ability to improve the spatial properties of a clinical treatment plan while preserving the target coverage and the dose to the organs at risk. The goal was to obtain a more homogenous treatment plan, minimizing the need for manual adjustments after inverse treatment planning.

Materials And Methods: The study included 25 previously treated prostate cancer patients.

View Article and Find Full Text PDF

Background: Spatial properties of a dose distribution, such as volumes of contiguous hot spots, are of clinical importance in treatment planning for high dose-rate brachytherapy (HDR BT). We have in an earlier study developed an optimization model that reduces the prevalence of contiguous hot spots by modifying a tentative treatment plan.

Purpose: The aim of this study is to incorporate the correction of hot spots in a standard inverse planning workflow and to validate the integrated model in a clinical treatment planning system.

View Article and Find Full Text PDF

Background: Intensity-modulated brachytherapy (IMBT) is an emerging technology for cancer treatment, in which radiation sources are shielded to shape the dose distribution. The rotatable shields provide an additional degree of freedom, but also introduce an additional, directional, type of uncertainty, compared to conventional high-dose-rate brachytherapy (HDR BT).

Purpose: We propose and evaluate a robust optimization approach to mitigate the effects of rotational uncertainty in the shields with respect to planning criteria.

View Article and Find Full Text PDF

The vast majority of radiotherapy departments in Europe using brachytherapy (BT) perform temporary implants of high- or pulsed-dose rate (HDR-PDR) sources with photon energies higher than 50 keV. Such techniques are successfully applied to diverse pathologies and clinical scenarios. These recommendations are the result of Working Package 21 (WP-21) initiated within the BRAchytherapy PHYsics Quality Assurance System (BRAPHYQS) GEC-ESTRO working group with a focus on HDR-PDR source calibration.

View Article and Find Full Text PDF

Background: There is increased interest in in vivo dosimetry for Ir brachytherapy (BT) treatments using high atomic number (Z) inorganic scintillators. Their high light output enables construction of small detectors with negligible stem effect and simple readout electronics. Experimental determination of absorbed-dose energy dependence of detectors relative to water is prevalent, but it can be prone to high detector positioning uncertainties and does not allow for decoupling of absorbed-dose energy dependence from other factors affecting detector response .

View Article and Find Full Text PDF

Brachytherapy is a mature treatment modality. The literature is abundant in terms of review articles and comprehensive books on the latest established as well as evolving clinical practices. The intent of this article is to part ways and look beyond the current state-of-the-art and review emerging technologies that are noteworthy and perhaps may drive the future innovations in the field.

View Article and Find Full Text PDF

Background And Purpose: Brachytherapy treatment outcomes depend on the accuracy of the delivered dose distribution, which is proportional to the reference air-kerma rate (). Current societal recommendations require the medical physicist to compare the measured values to the manufacturer source calibration certificate. The purpose of this work was to report agreement observed in current clinical practice in the European Union.

View Article and Find Full Text PDF

Dual-energy computed tomography (CT) can be used in radiotherapy treatment planning for the calculation of absorbed dose distributions. The aim of this work is to evaluate whether there is room for improvement in the accuracy of the Monoenergetic Plus algorithm by Siemens Healthineers. A Siemens SOMATOM Force scanner was used to scan a cylindrical polymethyl methacrylate phantom with four rod-inserts made of different materials.

View Article and Find Full Text PDF

The choice of the material base to which the material decomposition is performed in dual-energy computed tomography may affect the quality of reconstructed images. The aim of this work is to investigate how the commonly used bases (water, bone), (water, iodine) and (photoelectric effect, Compton scattering) affect the reconstructed linear attenuation coefficient in the case of the Alvarez-Macovski method. The performance of this method is also compared with the performance of the Dual-energy Iterative Reconstruction Algorithm (DIRA).

View Article and Find Full Text PDF

Cerium oxide nanoparticles with integrated gadolinium have been proved to be useful as contrast agents in magnetic resonance imaging. Of question is their performance in dual-energy computed tomography. The aims of this work are to determine (1) the relation between the computed tomography number and the concentration of the I, Gd or Ce contrast agent and (2) under what conditions it is possible to resolve the type of contrast agent.

View Article and Find Full Text PDF

Automatic segmentation of bones in computed tomography (CT) images is used for instance in beam hardening correction algorithms where it improves the accuracy of resulting CT numbers. Of special interest are pelvic bones, which-because of their strong attenuation-affect the accuracy of brachytherapy in this region. This work evaluated the performance of the JJ2016 algorithm with the performance of MK2014v2 and JS2018 algorithms; all these algorithms were developed by authors.

View Article and Find Full Text PDF

Purpose: High dose-rate prostate brachytherapy has been implemented in Sweden in the late 1980s and early 1990s in six clinics using the same schedule: 20 Gy in two fractions combined with 50 Gy in 25 fractions with external beam radiation therapy. Thirty years have passed and during these years, various aspects of the treatment process have developed, such as ultrasound-guided imaging and treatment planning system. An audit was conducted, including a questionnaire and treatment planning, which aimed to gather knowledge about treatment planning methods in Swedish clinics.

View Article and Find Full Text PDF

Diamond detectors are increasingly employed in dosimetry. Their response has been investigated by means of Monte Carlo (MC) methods, but there is no consensus on what mass density, mean excitation energyand number of conduction electrons per atomto use in the simulations. The ambiguity occurs due to its seeming similarity with graphite (both are carbon allotropes).

View Article and Find Full Text PDF

Treatment planning in high dose-rate brachytherapy has traditionally been conducted with manual forward planning, but inverse planning is today increasingly used in clinical practice. There is a large variety of proposed optimization models and algorithms to model and solve the treatment planning problem. Two major parts of inverse treatment planning for which mathematical optimization can be used are the decisions about catheter placement and dwell time distributions.

View Article and Find Full Text PDF

Existing methods for checking the light field-radiation field congruence on x-ray equipment either do not fully meet the conditions of various quality control standards regarding inherent uncertainty requirements or contain subjective steps, further increasing the uncertainty of the end result. The aim of this work was to develop a method to check the light field-radiation field congruence on all x-ray equipment. The result should have a low uncertainty which is accomplished by eliminating most subjective user steps in the method.

View Article and Find Full Text PDF

Optimising phosphor screens in dose detectors or imaging sensor designs is a cumbersome and time- consuming work normally involving specialised measuring equipment and advanced modelling. It is known that crucial optical parameters of the same phosphor may vary within a wide range of values. The aim of this work was to experimentally assess a simple previously published model where the case specific optical parameters (scattering and absorption) are instead represented by a fixed, single parameter, the light extinction factor, ξ.

View Article and Find Full Text PDF

Deep learning algorithms have improved the speed and quality of segmentation for certain tasks in medical imaging. The aim of this work is to design and evaluate an algorithm capable of segmenting bones in dual-energy CT data sets. A convolutional neural network based on the 3D U-Net architecture was implemented and evaluated using high tube voltage images, mixed images and dual-energy images from 30 patients.

View Article and Find Full Text PDF

Purpose: An important characteristic of radiation dosimetry detectors is their energy response which consists of absorbed-dose and intrinsic energy responses. The former can be characterized using Monte Carlo (MC) simulations, whereas the latter (i.e.

View Article and Find Full Text PDF

High dose-rate brachytherapy is a modality of radiation therapy used for cancer treatment, in which the radiation source is placed within the body. The treatment goal is to give a high enough dose to the tumour while sparing nearby healthy tissue and organs (organs-at-risk). The most common criteria for evaluating dose distributions are dosimetric indices.

View Article and Find Full Text PDF