This study evaluates the volumetric modulated arc therapy (VMAT) dosimetric comparison between Halcyon ring gantry and TrueBeam c-arm linear accelerators for craniospinal irradiation (CSI) of the neuro-axis. 25 patients, who received treatment for medulloblastoma and primitive neuro-ectodermal tumors between 2018 and 2021, were planned for VMAT in True Beam (TB), and Halcyon (HAL) linear accelerators using 6 MV unflattened (FFF) photon beams (HALFFF and TBFFF). Dose-volume statistics for the target and organs at risk (OARs) and the total number of monitoring units (MUs) in the treatment plans were compared which included dose received by 95% PTV volume (V95%), volume receiving ≥ 107% dose, homogeneity index (HI), conformity index (PI), MU and dose spillage (D10%, D30%, D50%, D70%, D90%).
View Article and Find Full Text PDFAim: This study aims to report preclinical validation, and the first clinical treatment of total bone marrow irradiation (TMI) and total bone marrow and lymph nodal irradiation (TMLI) using Volumetric modulated arc therapy in Halcyon-E ring gantry linear accelerator. Preclinical validation includes simulation, planning, patient-specific QA, and dry run.
Material And Method: Four patients, two female and two male, with body weights of 116 kg, 52 kg, 64 kg, and 62 kg; with two with chronic myeloid leukemia, one each with acute lymphoblastic leukemia and acute myeloid leukemia (AML) were simulated and planned for TMI/TMLI.