Purpose: The aim of this work was to evaluate the various computed tomography (CT) techniques such as fast CT, slow CT, breath-hold (BH) CT, full-fan cone beam CT (FF-CBCT), half-fan CBCT (HF-CBCT), and average CT for delineation of internal target volume (ITV). In addition, these ITVs were compared against four-dimensional CT (4DCT) ITVs.
Materials And Methods: Three-dimensional target motion was simulated using dynamic thorax phantom with target insert of diameter 3 cm for ten respiration data.
Tumors in thoracic and upper abdomen regions such as lungs, liver, pancreas, esophagus, and breast move due to respiration. Respiration-induced motion introduces uncertainties in radiotherapy treatments of these sites and is regarded as a significant bottleneck in achieving highly conformal dose distributions. Recent developments in radiation therapy have resulted in (i) motion-encompassing, (ii) respiratory gating, and (iii) tracking methods for adapting the radiation beam aperture to account for the respiration-induced target motion.
View Article and Find Full Text PDFPurpose: With the advent of state-of-the-art treatment technologies, the use of small fields has increased, and dosimetry in small fields is highly challenging. In this study, the potential use of Varian electronic portal imaging device (EPID) for small field measurements was explored for 6 and 15 MV photon beams.
Materials And Methods: The output factors and profiles were measured for a range of jaw-collimated square field sizes starting from 0.
Objective: The purpose of this work was to evaluate the four-dimensional cone beam CT (4DCBCT) imaging with different gantry rotation speed.
Methods: All the 4DCBCT image acquisitions were carried out in Elekta XVI Symmetry™ system (Elekta AB, Stockholm, Sweden). A dynamic thorax phantom with tumour mimicking inserts of diameter 1, 2 and 3 cm was programmed to simulate the respiratory motion (4 s) of the target.
Objective: The aim of this work was to evaluate the quality of kilovoltage (kV) cone beam CT (CBCT) images acquired during arc delivery.
Methods: Arc plans were delivered on a Catphan(®) 600 phantom (The Phantom Laboratory Inc., Salem, NY), and kV CBCT images were acquired during the treatment.
The purpose of this study was to evaluate the capabilities of DMLC to deliver the respiratory motion-synchronized dynamic IMRT (MS-IMRT) treatments under various dose rates. In order to create MS-IMRT plans, the DMLC leaf motions in dynamic IMRT plans of eight lung patients were synchronized with the respiratory motion of breathing period 4 sec and amplitude 2 cm (peak to peak) using an in-house developed leaf position modification program. The MS-IMRT plans were generated for the dose rates of 100 MU/min, 400 MU/min, and 600 MU/min.
View Article and Find Full Text PDFAustralas Phys Eng Sci Med
March 2012
The purpose of this study is to analyze the effect of various dose rates (DR) and maximum allowable MLC leaf velocities (MLV) in dynamic Intensity Modulated Radiotherapy (IMRT) planning and delivery of head and neck patients. Five head and neck patients were retrospectively included in this study. The initial dynamic IMRT 'reference plans' were created for all these patients, using a DR of 400 MU/min and MLV of 2.
View Article and Find Full Text PDFThe purpose of this study was to evaluate the dosimetric effect of the leaf width of a multileaf collimator (MLC) in intensity-modulated radiotherapy (IMRT) delivery techniques for small- and large-volume targets. We retrospectively selected previously treated 5 intracranial and 5 head-neck patients for this study to represent small- (range, 18.37-72.