The two-dimensional lateral dose profiles D(x, y) of narrow photon beams, typically used for beamlet-based IMRT, stereotactic radiosurgery and tomotherapy, can be regarded as resulting from the convolution of a two-dimensional rectangular function R(x, y), which represents the photon fluence profile within the field borders, with a rotation-symmetric convolution kernel K(r). This kernel accounts not only for the lateral transport of secondary electrons and small-angle scattered photons in the absorber, but also for the 'geometrical spread' of each pencil beam due to the phase-space distribution of the photon source. The present investigation of the convolution kernel was based on an experimental study of the associated line-spread function K(x).
View Article and Find Full Text PDFThe unshielded Si diode PTW 60012, used for accurate measurements of the transversal dose profiles of narrow photon beams, has been investigated with regard to its linearity, photon energy dependence and spatial resolution. The diode shows a slight supralinearity, i.e.
View Article and Find Full Text PDFThe spatial resolution of 2D detector arrays equipped with ionization chambers or diodes, used for the dose verification of IMRT treatment plans, is limited by the size of the single detector and the centre-to-centre distance between the detectors. Optimization criteria with regard to these parameters have been developed by combining concepts of dosimetry and pattern analysis. The 2D-ARRAY Type 10024 (PTW-Freiburg, Germany), single-chamber cross section 5 x 5 mm(2), centre-to-centre distance between chambers in each row and column 10 mm, served as an example.
View Article and Find Full Text PDFPurpose: To investigate the attenuation of a carbon-fiber tabletop and a combiboard, alongside with the depth-dose profile in a solid-water phantom.
Material And Methods: Depth-dose measurements were performed with a Roos chamber for 6- and 10-MV beams for a typical field size (15 cm x 15 cm, SSD [source-surface distance] 100 cm). A rigid-stem ionization chamber was used to measure transmission factors.
In the dosimetry of narrow photon fields with side lengths of the order of 1 cm, the traditional parametrisation via the absolute dose on the beam axis and the relative lateral dose distribution has to deal with the difficulty to find sufficiently small detectors and to adjust them accurately on the narrow-beam axis. This can be avoided by reconsidering the parametrisation, using as normalization factor the surface integral of the dose in the plane perpendicular to the beam axis, abbreviated as the "dose-area product" (DAP). We investigated and confirmed the ability of a large-area parallel-plate ionisation chamber, with a sensitive volume shaped as a flat cylinder of 81.
View Article and Find Full Text PDFIn this paper we describe a concept for dosimetric treatment plan verification using two-dimensional ionization chamber arrays. Two different versions of the 2D-ARRAY (PTW-Freiburg, Germany) will be presented, a matrix of 16 x 16 chambers (chamber cross section 8 mm x 8 mm; the distance between chamber centers, 16 mm) and a matrix of 27 x 27 chambers (chamber cross section 5 mm x 5 mm; the distance between chamber centers is 10 mm). The two-dimensional response function of a single chamber is experimentally determined by scanning it with a slit beam.
View Article and Find Full Text PDFThe two-dimensional verification of intensity-modulated radiation plans is one of the major requirements for the safe application of this technique. The present study examines the resolution and sensitivity of a two-dimensional ionisation-chamber array (PTW2D-Array, type 10024), which can be used for plan verification instead of films. According to the Shannon-Nyquist theorem, the resolution of the 2D-Array is sufficient for dose distributions with a minimal field size of 2 cm x 2 cm.
View Article and Find Full Text PDF