Diagnostic radiology dosimeters should comply with International Electrotechnical Commission (IEC) 61674 standard in order to perform measurements with sufficient accuracy and reliability. The calibration of a dosimeter is performed under, and pertains to, reference conditions. However, in most cases, dosimeters are used for clinical measurements under non-reference conditions.
View Article and Find Full Text PDFObjectives: (a) To measure the absorbed radiation doses at 16 anatomical sites of a Rando phantom and (b) to calculate the effective doses including and excluding the salivary gland doses in panoramic radiography using a conventional and a digital panoramic device.
Study Design: Thermoluminescent dosimeters (TLD-100) were placed at 16 sites in a Rando phantom, using a conventional, Planmeca Promax and a digital, Planmeca PM2002CC Proline 2000 (Planmeca Oy, 00880 Helsinki, Finland) panoramic device for panoramic radiography. During conventional radiography the selected exposure settings were 66 kVp, 6 mA and 16s, while during digital radiography two combinations were selected 60 kVp, 4 mA, 18 s and 66 kVp, 8 mA, 18s with and without image processing function.