This was a head-to-head comparative study on different electrocardiogram (ECG)-based smartwatches and devices for atrial fibrillation detection. We prospectively included 220 patients scheduled for electrical cardioversion and recorded ECGs with 3 different devices (Withings Move ECG, Apple Watch 5, Kardia Mobile 6-leads) as well as the standard 12-lead ECG (gold standard), both before and after cardioversion. All atrial fibrillation detection algorithms had high accuracy (sensitivity and specificity: 91-99%) but were hampered by uninterpretable recordings (20-24%).
View Article and Find Full Text PDF