  • A new dual chamber leadless pacemaker system utilizes a temperature-based sensor to monitor heart rate response during exercise, focusing on its effectiveness in both the right ventricle and right atrium.
  • An exercise test was conducted 6 weeks post-implant on eligible subjects to evaluate the pacing response of the ventricular and atrial devices, measuring if the rate increased proportionately with metabolic workload.
  • Results showed that both the ventricular and atrial devices had mean rate-response slopes (0.93 and 0.91, respectively) within the acceptable range, indicating that the temperature-based sensor effectively adjusts pacing to meet increased metabolic demands.