Purpose: Multichannel Transcranial Magnetic Stimulation (mTMS) arrays enable multiple sites to be stimulated simultaneously or sequentially under electronic control without moving the system's stimulation coils. Here, we build and characterize the performance of a novel modular 3-axis TMS coil that can be utilized as a unit element in large-scale multichannel TMS arrays.
Methods: We determined the basic physical characteristics of the 3-axis TMS coil x-, y- and z-elements using a custom 2-channel programmable stimulator prototype.
We developed a 375-channel, whole-head magnetoencephalography (MEG) system ("BabyMEG") for studying the electrophysiological development of human brain during the first years of life. The helmet accommodates heads up to 95% of 36-month old boys in the USA. The unique two-layer sensor array consists of: (1) 270 magnetometers (10 mm diameter, ∼15 mm coil-to-coil spacing) in the inner layer, (2) thirty-five three-axis magnetometers (20 mm × 20 mm) in the outer layer 4 cm away from the inner layer.
View Article and Find Full Text PDFBackground: A major motivation in designing the new infant and child magnetoencephalography (MEG) system described in this manuscript is the premise that electrophysiological signatures (resting activity and evoked responses) may serve as biomarkers of neurodevelopmental disorders, with neuronal abnormalities in conditions such as autism spectrum disorder (ASD) potentially detectable early in development. Whole-head MEG systems are generally optimized/sized for adults. Since magnetic field produced by neuronal currents decreases as a function of distance(2) and infants and young children have smaller head sizes (and thus increased brain-to-sensor distance), whole-head adult MEG systems do not provide optimal signal-to-noise in younger individuals.
View Article and Find Full Text PDF