Background And Purpose: Adding hydroxyapatite to a stem to enhance ingrowth is a matter of debate, even less is known about the long-term effect on stability by adding hydroxyapatite (HA). Continuous migration in the first 2-5 years is an indicator of failed osteointegration or pending failure, enhancing the risk of loosening within 10 years after initial surgery. We performed a double-blinded randomised RSA trial with 5-year follow-up, to compare and analyse migration characteristics of the hydroxyapatite uncoated (HA-) and hydroxyapatite coated (HA+) Zweymuller-type hip stem.