In order to assess SARS-CoV-2 real time quantitative polymerase chain reaction (RT-qPCR) results in a real-life setting, three independent laboratories in Graz (Austria) set up a continuous cross comparison schedule. The following test systems were used: The QIAGEN NeuMoDx SARS-CoV-2 Assay, the Allplex™ 2019-nCoV Assay (Seegene) on a MicroLab Nimbus (Hamilton) platform combined with RealStar SARS-CoV-2 RT-PCR Assay (Altona Diagnostics GmbH), and the cobas SARS-CoV-2 test on a fully automated cobas 6800 system (Roche). A total of 200 samples were analysed, 184 (92%) were found to be concordant with all testing platforms, 14 (7%) discordant.