Background: Using middleware solutions, it is possible to implement concentration-dependent analyte-specific hemolysis rejection limits. This makes day-to-day reporting of clinical specimens more efficient and potentially lowers sample rejection rates compared to a "one-size-fits-all" approach (i.e.
View Article and Find Full Text PDFObjectives: We sought to replace the highly hemolysis-susceptible diazo conjugated bilirubin (Bc) assay with the more robust vanadate oxidation method and determine its impact on test cancellation in the pediatric population.
Methods: Analytical validation of the Randox vanadate assay and comparison with the Roche diazo method were performed. The frequency of pediatric sample cancellation because of hemolysis was compared between the diazo and vanadate methods by retrospective analysis of clinical test data.
Objectives: To analytically and clinically evaluate the semiquantitative Elecsys anti-severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) spike protein antibody (S-Ab) assay on the Roche cobas e602 analyzer.
Methods: The S-Ab assay is a 1-step, double-antigen sandwich electrochemiluminescent immunoassay that semiquantitatively measures total IgG, IgM, and IgA antibodies specific for the receptor binding domain of SARS-CoV-2 spike protein in serum or plasma. The S-Ab assay was evaluated for precision, linearity, interference (by hemoglobin, bilirubin, triglycerides, and biotin), cross-reactivity, and clinical performance, and was compared to the qualitative Elecsys anti-nucleocapsid (N-Ab) immunoassay, a lateral flow device that qualitatively detects S-Ab and N-Ab, and an anti-spike enzyme-linked immunosorbent assay (ELISA).