A new calibration methodology for regenerated fiber Bragg grating (RFBG) temperature sensors up to 700 °C is proposed and demonstrated. A generalized, wavelength-dependent temperature calibration function is experimentally determined that describes the temperature-induced wavelength shifts for all RFBG sensor elements that are manufactured with the same fabrication parameters in the wavelength range from 1465 nm to 1605 nm. Using this generalized calibration function for absolute temperature measurements, each RFBG sensor element only needs to be calibrated at one reference temperature, representing a considerable simplification of the conventional calibration procedure.