Purpose: Standardized bench testing of the new ShockPulse™ intracorporeal lithotripter was performed against three commercially available lithotripsy systems to determine differences and nuances in performance.

Materials And Methods: The ShockPulse intracorporeal lithotripter was tested against the LUS-2™, CyberWand,™ and EMS LithoClast™ in a standardized bench setting using hard (Ultracal-30) and soft (plaster of Paris) stone phantoms. An in vitro kidney model was used to record the time needed to fragment stone samples into retrievable-sized pieces.