Publications by authors named "Anant Mandapaka"

Purpose: The continuous scanning mode of electronic portal imaging devices (EPID) that offers time-resolved information has been newly explored for verifying dynamic radiation deliveries. This study seeks to determine operating conditions (dose rate stability and time resolution) under which that mode can be used accurately for the time-resolved dosimetry of intensity-modulated radiation therapy (IMRT) beams.

Methods: The authors have designed the following test beams with variable beam holdoffs and dose rate regulations: a 10 × 10 cm open beam to serve as a reference beam; a sliding window (SW) beam utilizing the motion of a pair of multileaf collimator (MLC) leaves outside the 10 × 10 cm jaw; a step and shoot (SS) beam to move the pair in step; a volumetric modulated arc therapy (VMAT) beam.

View Article and Find Full Text PDF

A novel tissue-equivalent proportional counter (TEPC) based on a gas electron multiplier (GEM) for measuring H*(10) for neutrons was designed and constructed. The pulse height spectra (PHS) of two different neutron sources (a 252Cf source and a AmBe source) were measured using the new TEPC. The measurements were made with the TEPC filled with two different gases (10P gas and a propane-based tissue-equivalent gas) at various pressures.

View Article and Find Full Text PDF