Objectives: Analytical validation of automated erythrocyte sedimentation rate (ESR) analyzers is necessary prior to their implementation into routine practice. Our aim was to perform the analytical validation of the modified Westergren method applied on the CUBE 30 touch analyzer (Diesse, Siena, Italy).
Methods: Validation included determination of within-run and between-run precision following the Clinical and Laboratory Standards Institute EP15-A3 protocol, comparison with the reference Westergren method, sample stability assessment at both room temperature and 4 °C, after 4, 8 and 24-h storage, and checking the extent of hemolysis and lipemia interference.
Objectives: The body of literature varies significantly regarding serum and urine osmolality stability. Therefore, our aim was to investigate the stability of serum and urine osmolality at different temperatures (room temperature (RT) 4-8 °C, -20 °C) and time conditions (8 h, 24 h, 1 month).
Methods: The stability study was conducted following the CRESS guidelines, including 40 serum and urine samples.
Introduction: Serum samples of haemodialysed patients collected through vascular access devices, central venous catheter (CVC) can contain residual heparin, which can cause incomplete clotting and consequently fibrinogen interference in serum protein electrophoresis (SPE). We hypothesized that this problem may be overcome by addition of thrombin and aimed to find a simple thrombin-based method for fibrinogen interference removal.
Materials And Methods: Blood samples of 51 haemodialysed patients with CVC were drawn through catheter into Clot Activator Tube (CAT) and Rapid Serum Tube Thrombin (RST) vacutainers (Becton Dickinson, New Jersey, USA) following the routine hospital protocols and analysed with gel-electrophoresis (Sebia, Lisses, France).