Comput Methods Biomech Biomed Engin
November 2015
Veneer fracture is the most common complication in zirconia-based restorations. The aim of this study was to evaluate the mechanical behavior of a zirconia-based crown in a lower canine tooth supporting removable partial denture (RPD) prosthesis, varying the bond quality of the veneer/coping interface. Microtomography (μCT) data of an extracted left lower canine were used to build the finite element model (M) varying the core material (gold core - MAu; zirconia core - MZi) and the quality of the veneer/core interface (complete bonded - MZi; incomplete bonded - MZi-NL).
View Article and Find Full Text PDFPurpose: To evaluate and compare the reliability of implant-supported single crowns cemented onto abutments retained with coated (C) or noncoated (NC) screws and onto platform-switched abutments with coated screws.
Materials And Methods: Fifty-four implants (DT Implant 4-mm Standard Platform, Intra-Lock International) were divided into three groups (n = 18 each) as follows: matching-platform abutments secured with noncoated abutment screws (MNC); matching-platform abutments tightened with coated abutment screws (MC); and switched-platform abutments secured with coated abutment screws (SC). Screws were characterized by scanning electron microscopy and x-ray photoelectron spectroscopy (XPS).
Int J Oral Maxillofac Implants
February 2014
Purpose: This study evaluated the reliability and failure modes of anterior implants with internal-hexagon (IH), external-hexagon (EH), or Morse taper (MT) implant-abutment interface designs. The postulated hypothesis was that the different implant-abutment connections would result in different reliability and failure modes when subjected to step-stress accelerated life testing (SSALT) in water.
Materials And Methods: Sixty-three dental implants (4 × 10 mm) were divided into three groups (n = 21 each) according to connection type: EH, IH, or MT.
Purpose: To evaluate the stress distribution in peri-implant bone by simulating the effect of an implant with microthreads and platform switching on angled abutments through tridimensional finite element analysis. The postulated hypothesis was that the presence of microthreads and platform switching would reduce the stress concentration in the cortical bone.
Methods: Four mathematical models of a central incisor supported by an implant (5.
Purpose: The aim of this study was to evaluate stress distribution on peri-implant bone simulating the influence of implants with different lengths on regular and switching platforms in the anterior maxilla by means of three-dimensional finite element analysis.
Materials And Methods: Four mathematical models of a central incisor supported by an external hexagon implant (diameter, 5.0 mm) were created, varying the length (15.
Purpose: In view of reports in the literature on the benefits achieved with the use of platform switching, described as the use of an implant with a larger diameter than the abutment diameter, the goal being to prevent the (previously) normal bone loss down to the first thread that occurs around most implants, thus enhancing soft tissue aesthetics and stability and the need for implant inclination due to bone anatomy in some cases, the aim of this study was to evaluate bone stress distribution on peri-implant bone, by using three-dimensional finite element analysis to simulate the influence of implants with different abutment angulations (0 and 15 degrees) in platform switching.
Methods: Four mathematical models of an implant-supported central incisor were created with varying abutment angulations: straight abutment (S1 and S2) and angulated abutment at 15 degrees (A1 and A2), submitted to 2 loading conditions (100 N): S1 and A1-oblique loading (45 degrees) and S2 and A2-axial loading, parallel to the long axis of the implant. Maximum (σmax) and minimum (σmin) principal stress values were obtained for cortical and trabecular bone.
Purpose: This study used the 3D finite element (FE) method to evaluate the mechanical behavior of a maxillary central incisor with three types of dowels with variable heights of the remaining crown structure, namely 0, 1, and 2 mm.
Materials And Methods: Based on computed microtomography, nine models of a maxillary central incisor restored with complete ceramic crowns were obtained, with three ferrule heights (0, 1, and 2 mm) and three types of dowels (glass fiber = GFD; nickel-chromium = NiCr; gold alloy = Au), as follows: GFD0--restored with GFD with absence (0 mm) of ferrule; GFD1--similar, with 1 mm ferrule; GFD2--glass fiber with 2 mm ferrule; NiCr0--restored with NiCr alloy dowel with absence (0 mm) of ferrule; NiCr1--similar, with 1 mm ferrule; NiCr2--similar, with 2 mm ferrule; Au0--restored with Au alloy dowel with absence (0 mm) of ferrule; Au1--similar, with 1 mm ferrule; Au2--similar, with 2 mm ferrule. A 180 N distributed load was applied to the lingual aspect of the tooth, at 45° to the tooth long axis.
Purpose: This study aimed to evaluate stress distribution on peri-implant bone simulating the influence of platform switching in external and internal hexagon implants using three-dimensional finite element analysis.
Materials And Methods: Four mathematical models of a central incisor supported by an implant were created: External Regular model (ER) with 5.0 mm × 11.
The aim of this study was to describe the treatment used in an elderly patient presenting with bruxism and dental erosion, with good gingival health and bone support, but with decreased occlusal vertical dimension (OVD). The oral rehabilitation of elderly patients presenting with bruxism in association with tooth erosion has been a great challenge for dentists. The loss of OVD, the presence of occlusal instability and the absence of an effective anterior guide due excessive dental wear, can damage stomatognathic system (SS) biology, the function and the aesthetics.
View Article and Find Full Text PDFObjectives: Because the mechanical behavior of the implant-abutment system is critical for the longevity of implant-supported reconstructions, this study evaluated the fatigue reliability of different implant-abutment systems used as single-unit crowns and their failure modes.
Methods And Materials: Sixty-three Ti-6Al-4V implants were divided in 3 groups: Replace Select (RS); IC-IMP Osseotite; and Unitite were restored with their respective abutments. Anatomically correct central incisor metal crowns were cemented and subjected to separate single load to failure tests and step-stress accelerated life testing (n = 18).
Purpose: Failures of implant-abutment connections have been observed clinically, especially in single-tooth replacements. This study sought to evaluate the reliability and failure modes of implant-supported anterior crowns restored with different implant systems.
Materials And Methods: Forty-two Ti-6Al-4V dental implants (~4 mm diameter) were used for single anterior crown replacement and divided into two groups according to tested system: (NB) Replace Select system, Nobel Biocare (n = 21); and (IL) Internal connection system, Intra-Lock International (n = 21).
The mechanical performance of cemented or screw-retained implant-supported crowns with an internal or external configuration is yet to be understood. This in vitro study evaluated the effect of screw-retained and cement-retained prostheses on internal and external implant-abutment connections. Thereby, the reliability and failure modes of crowns were investigated.
View Article and Find Full Text PDFObjectives: Based on a maxillary premolar restored with laminate veneer and using the 3-D finite element analysis (FEA) and μCT data, the aim of this study was to evaluate the influence of different types of buccal cusp reduction on the stress distribution in the porcelain laminate veneer and in the resin luting cement layer.
Methods: Two 3-D FEA models (M) of a maxillary premolar were built from μCT data. The buccal cusp reduction followed two configurations: Mt - buccal cusp completely covered by porcelain laminate veneer; and Mp - buccal cusp partially covered by porcelain laminate veneer.
Clin Oral Implants Res
January 2012
Objectives: To evaluate the effect of insertion torque on micromotion to a lateral force in three different implant designs.
Material And Methods: Thirty-six implants with identical thread design, but different cutting groove design were divided in three groups: (1) non-fluted (no cutting groove, solid screw-form); (2) fluted (90° cut at the apex, tap design); and (3) Blossom(™) (Patent pending) (non-fluted with engineered trimmed thread design). The implants were screwed into polyurethane foam blocks and the insertion torque was recorded after each turn of 90° by a digital torque gauge.
Purpose: The purpose of this study was to evaluate the influence of different types of bone on the stress distribution in the mandibular bone supporting a prefabricated bar-type implant prosthesis using three-dimensional finite element analysis.
Materials And Methods: Four finite element models (M) of a completely edentulous mandibular arch were built. The bone types varied from type 1 to type 4 (M1, M2, M3, M4).
Statement Of Problem: The fracture or chipping of ceramic veneers is a common problem for zirconia-based restorations.
Purpose: This study evaluated the stress distribution in the veneer of a maxillary central incisor restored with a complete crown using a zirconia core with a feldspathic ceramic veneer, simulating an incomplete bond between the veneer and zirconia substructure.
Material And Methods: Based on a microcomputed tomography of a maxillary central incisor, 3 finite element models (M) for a complete crown were developed: Mf, a complete crown based on feldspathic ceramic; Mlz, a zirconia-based complete crown with a complete bond at the zirconia/veneer interface; and Mnzl, similar to Mlz, but with an incomplete bond at the zirconia/veneer interface created by using a contact element with a frictional coefficient of 0.
The aim of the present study was to conduct a critical literature review about the technique of computer-guided surgery in implantology to highlight the indications, purposes, immediate loading of implants and complications, protocol of fabrication, and functioning of virtual planning software. This literature review was based on OLDMEDLINE and MEDLINE databases from 2002 to 2010 using the key words "computer-guided surgery" and "implant-supported prosthesis." Thirty-four studies regarding this topic were found.
View Article and Find Full Text PDFPurpose: The aim of this study was to evaluate the influence of different Young moduli of the ceramic crown on the distribution of tensions in the region of the abutment-crown interface by making use of 2D finite element analysis.
Materials: Two representative models of a sagittally sectioned maxilla were built through AutoCad program showing an implant in the region of the upper central incisor and were restored by means of IPS e.max Press or Procera AllCeram on zircon abutment.
Comput Methods Biomech Biomed Engin
October 2010
This study aimed to evaluate the influence of loading on a maxillary central incisor with the periodontal ligament (PDL) represented by 2D elastic beam elements using a 2D finite element analysis. Two models (M) were built varying the PDL representation: Mh (homogeneous PDL) and Mht (heterogeneous PDL with beam3 elements). Stress and displacements were determined for three loading conditions (L): Ll, lingual face loading at 45° with the tooth long axis; Li, perpendicular to the incisal edge; and Lip, on the incisal edge, parallel to the tooth long axis.
View Article and Find Full Text PDF