In the presence of haemolysis, the interpretation of the Lactate dehydrogenase (LDH) activity result is a major operational challenge for medical laboratories: if the origin is intravascular, then the measurement will reflect the clinical reality, but in extravascular haemolysis, the laboratory will be confronted with an artefactual increase leading to false-positive high results. The aim of our study was to evaluate the adjustment of LDH concentration results according to the haemolysis index (HI). After designed a mathematical model to correct the LDH measured as a function of the haemolysis index using a Cobas 8000 analyser (Roche diagnostics, Mannheim, Germany), LDH measurement of seventy-four duplicate samples were tested before and after exposure to extravascular haemolysis process.
View Article and Find Full Text PDFRoche Diagnostics decided a more restrictive haemolysis index (≤20 HI) for approval the release of serum potassium results. This study examined the risk of overestimating serum potassium results related to the HI according to the Ricos total error (±5.6%) and evaluated the approach of Martinez-Morillo and Alvarez to determine a corrected potassium.
View Article and Find Full Text PDF