Objective: This study set out to determine the accuracy with which tone pip ABR and click ABR, carried out in babies referred from universal newborn hearing screening, is able to predict the hearing outcome as determined by follow-up hearing tests.
Study Sample: The cohort of babies studied were all babies referred for hearing assessment from the universal newborn hearing screen in Sheffield, UK for the period January 2002 to September 2007, who were found to have a significant hearing impairment.
Design: The results of hearing assessment following referral from the newborn hearing screen were collected together with those of follow-up tests carried out up to an age when behavioural testing had established ear- and frequency-specific thresholds at 0.
Spontaneous otoacoustic emissions (SOAEs) and synchronized spontaneous otoacoustic emissions (SSOAEs) were recorded using both the standard closed-canal method of recording and a novel open-canal method which involved suspending the probe at the entrance to the ear canal with no occluding tip. In both conditions, a probe tube microphone was inserted down the ear canal to measure the acoustic pressure near the tympanic membrane. Open- and closed-canal recordings were obtained in twelve otologically normal ears, all of which exhibited SSOAEs, and 6 of which exhibited SOAEs.
View Article and Find Full Text PDF