  • * Using a micromanipulator and a user-friendly interface, we achieved high precision in SI, with a success rate of 95.0% in tests on ex-vivo porcine eyes.
  • * Our modifications allow for the transformation of standard 40G SI cannulas into guided ones, paving the way for accurate delivery of innovative treatments for retinal diseases.
View Article and Find Full Text PDF

Purpose: The occurrence of iatrogenic retinal breaks (RB) in pars plana vitrectomy (PPV) is a complication that compromises the overall efficacy of the surgery. A subset of iatrogenic RB occurs when the retina (rather than the vitreous gel) is cut accidentally by the vitrector. We developed a smart vitrector that can detect in real-time potential iatrogenic RB and activate promptly a PPV machine response to prevent them.

View Article and Find Full Text PDF

We introduced and validated a method to miniaturize graded-index (GRIN) fiber-based optical coherence tomography (OCT) probes down to 70  μm in diameter. The probes consist in an assembly of single-mode (SM), coreless (CL), and graded-index (GRIN) fibers. We opted for a probe design enabling controlled size reduction by hydrogen fluoride etching.