Purpose To evaluate the correlation between ultrasonographic (US) point shear-wave elastography (SWE) and magnetic resonance (MR) elastography liver shear-wave speed (SWS) measurements in a pediatric population and to determine if US data dispersion affects this relationship. Materials and Methods Institutional review board approval was obtained for this HIPAA-compliant investigation; informed consent and patient assent (as indicated) were obtained. Patients (age range, 0-21 years) undergoing clinical liver MR elastography between July 2014 and November 2015 were prospectively enrolled.
View Article and Find Full Text PDFPurpose To prospectively assess agreement and repeatability of magnetic resonance (MR) elastography liver stiffness measurements across imager manufacturers, field strengths, and pulse sequences. Materials and Methods This prospective cross-sectional study was approved by the institutional review board; informed consent was obtained from all subjects. On the basis of an a priori power calculation, 24 volunteer adult subjects underwent MR elastography with four MR imaging systems (two vendors) and multiple pulse sequences (two-dimensional [2D] gradient-echo [GRE] imaging, 2D spin-echo [SE] echo-planar imaging, and three-dimensional [3D] SE echo-planar imaging).