Publications by authors named "Ahmed A Mohymen"

Using National Electrical Manufacturers Association (NEMA) phantom, the behavior of four distinct Positron Emission Tomography/Computed Tomography (PET/CT) reconstruction algorithms was investigated. These reconstruction algorithms were (Ordered Subset Expectation Maximization (OSEM), OSEM+ (Point Spread Function) PSF, OSEM + Time of Flight (TOF), and OSEM + TOF + PSF), and the focus was on sphere sizes and SBRs using recovery coefficients as a quantitation method. The obtained results demonstrated the significant effect of TOF on Gibbs artifact and Partial Volume Effect (PVE) at various Sphere-to-Background Ratios (SBRs).

View Article and Find Full Text PDF

Aim: This study aims to investigate the influence of voxel size and postfiltering on the quantification of standardized uptake value (SUV) in positron emission tomography/computed tomography (PET/CT) images.

Materials And Methods: National Electrical Manufacturers Association phantom with the spheres of different sizes were utilized to simulate the lesions. The phantom was scanned using a PET/CT scanner, and the acquired images were reconstructed using two different matrix sizes, (192 × 192) and (256 × 256), and a wide range of postfiltering values.

View Article and Find Full Text PDF

A PHP Error was encountered

Severity: Warning

Message: fopen(/var/lib/php/sessions/ci_session3b7srhnsgav1ou1fti98b04ti6nio4dt): Failed to open stream: No space left on device

Filename: drivers/Session_files_driver.php

Line Number: 177

Backtrace:

File: /var/www/html/index.php
Line: 316
Function: require_once

A PHP Error was encountered

Severity: Warning

Message: session_start(): Failed to read session data: user (path: /var/lib/php/sessions)

Filename: Session/Session.php

Line Number: 137

Backtrace:

File: /var/www/html/index.php
Line: 316
Function: require_once