1. Background & Purpose: Investigate the applicability of a series of detectors in small field dosimetry and the possible differences between their responses to FF and FFF beams. This work extends upon the series of detectors used by other authors to also include metal-oxide-semiconductor field-effect transistors (MOSFETs) detectors and radiochromic film.
View Article and Find Full Text PDFAim: To estimate angular response deviation of MOSFETs in the realm of intraoperative electron radiotherapy (IOERT), review their energy dependence, and propose unambiguous names for detector rotations.
Background: MOSFETs have been used in IOERT. Movement of the detector, namely rotations, can spoil results.
Purpose: To assess the electron beam monitoring statistical process control (SPC) in linear accelerator (linac) daily quality control. We present a long-term record of our measurements and evaluate which SPC-led conditions are feasible for maintaining control.
Methods: We retrieved our linac beam calibration, symmetry, and flatness daily records for all electron beam energies from January 2008 to December 2013, and retrospectively studied how SPC could have been applied and which of its features could be used in the future.
In vivo dosimetry is recommended in intraoperative electron radiotherapy (IOERT). To perform real-time treatment monitoring, action levels (ALs) have to be calculated. Empirical approaches based on observation of samples have been reported previously, however, our aim is to present a predictive model for calculating ALs and to verify their validity with our experimental data.
View Article and Find Full Text PDFBackground And Purpose: Industrial companies use failure mode and effect analysis (FMEA) to improve quality. Our objective was to describe an FMEA and subsequent interventions for an automated intraoperative electron radiotherapy (IOERT) procedure with computed tomography simulation, pre-planning, and a fixed conventional linear accelerator.
Material And Methods: A process map, an FMEA, and a fault tree analysis are reported.
Introduction: In vivo dosimetry is desirable for the verification, recording, and eventual correction of treatment in intraoperative electron radiotherapy (IOERT). Our aim is to share our experience of metal oxide semiconductor field-effect transistors (MOSFETs) and radiochromic films with patients undergoing IOERT using a general-purpose linac.
Materials And Methods: We used MOSFETs inserted into sterile bronchus catheters and radiochromic films that were cut, digitized, and sterilized by means of gas plasma.
Objective: Show that verification through cone beam Kv CT (CBKvCT) in a series of patients treated with 3D external radiotherapy (3DRT) for prostate cancer (PC) is related to a reduction in acute and late toxicity levels.
Materials And Method: A retrospective, non-randomized study of two homogeneous groups of patients treated between 2005 and 2008, 46 were verified using electronic portal devices (EPIDs) and 48 through CBKvCT. They received 3DRT for localized PC (T1-T3N0M0) and were prescribed the same doses.
We present a clinical case of a patient diagnosed with a retroperitoneal sarcoma, which received preoperative treatment with daily verification via computed tomography obtained with kilovoltage cone beam. We compare the benefit of this treatment compared to other conventional treatment without image guiding, reporting quantitative results.
View Article and Find Full Text PDF