Publications by authors named "A Wenman"

Introduction: The 10 Kilovoltage (kV) rule was a historic exposure adaption technique designed for film screen X-ray imaging to reduce ionising radiation dose without loss of image quality. This study evaluates knee X-ray radiation dose and image quality between standard patient exposure factors, the historic 10 kV rule (-50 % Milliampere-second (mAs), and a modified 10 kV rule (-75 % mAs) using a digital radiography (DR) system.

Method: Applying the exposure factors of 63 kV and 8 mAs (standard pre-set exposure), 73 kV and 4 mAs (historic 10 kV rule) and 73 kV and 2 mAs (modified 10 kV) to a phantom knee and recording entrance skin dose (ESD) using thermoluminescence dosemeters (TLDs).

View Article and Find Full Text PDF