Two absolute methods were used for standardizing the 65Zn solution received within the framework of the 2002-2003 full-scale BIPM comparison. The first method was the 4piPC-gamma-coincidence with efficiency extrapolation. The measurement conditions were adjusted so that the slope of the extrapolation curve was absent, which ensured that the measurement result did not depend on the positron branch intensity.
View Article and Find Full Text PDFA joint project has been established between VNIIM (D.I. Mendeleyev Institute for Metrology) and LNHB (Laboratoire National Henri Becquerel) to determine as accurately as possible the X- and gamma-ray emission probabilities of 154Eu.
View Article and Find Full Text PDFD.I. Mendeleyev Institute for Metrology (VNIIM) participated in two international comparisons of 204Tl and 89Sr solutions organized by the BIPM.
View Article and Find Full Text PDFWithin the framework of EUROMET project No. 410, standardization of a 169Yb solution has been carried out and the gamma-ray emission probabilities at its decay have been measured. 169Yb activity was measured by the 4pibeta-gamma-coincidence extrapolation method.
View Article and Find Full Text PDFThe half-life of 124Sb has been measured using two solutions prepared from 124Sb obtained from the nuclear reactions: 124Sn(p,n) 124Sb (cyclotron) and 123Sb(n,gamma) 124Sb (reactor). These solutions were standardized by the 4pibeta-gamma-coincidence extrapolation method with a total uncertainty of 0.12%.
View Article and Find Full Text PDF