Purpose: The purpose of study is to investigate the dosimetry of electron intraoperative radiotherapy (IOERT) of the Intraop Mobetron 2000 mobile LINAC in treatments outside of the breast. After commissioning and external validation of dosimetry, we report in vivo results of measurements for treatments outside the breast in a large patient cohort, and investigate if the presence of inhomogeneities can affect in vivo measurements.
Methods And Materials: Applicator factors and profile curves were measured with a stereotactic diode.
Introduction: This study aimed to assess the safety of high doses of radiation delivered with tomotherapy to the intact lung after radical pleurectomy/decortication or biopsy for malignant pleural mesothelioma (MPM).
Methods: Twenty-eight patients were enrolled in this prospective study and underwent adjuvant or definitive tomotherapy after radical pleurectomy/decortication (n = 20) or pleural biopsy (n = 8) for MPM. The dose prescribed to the planning target volume, defined as the entire hemithorax, including chest-wall incisions and drain sites and excluding the intact lung, was 50 Gy delivered in 25 fractions.
Purpose: EBT2 radiochromic films were studied and used for in vivo dosimetry in targeted intraoperative radiotherapy (TARGIT), a technique in which the Intrabeam system (Carl Zeiss, Oberkochen, Germany) is used to perform intraoperative partial breast irradiation with x-rays of 50 kV(p).
Methods: The energy of the radiation emitted by the Intrabeam with the different spherical applicators, under 1 and 2 cm of solid water, and under the tungsten impregnated rubber used for shielding of the heart in TARGIT of the breast, was characterized with measurements of half-value layer (HVL). The stability of response of EBT2 was verified inside this range of energies.
Helical Tomotherapy (HT) is a highly conformal image-guided radiation technique, introduced into clinical routine in 2006 at the Centro di Riferimento Oncologico Aviano (Italy). With this new technology, intensity-modulated radiotherapy (IMRT) is delivered using a helicoidal method. Here we present our dosimetric experiences using HT in 100 children, adolescents and young adults treated from May 2006 to February 2011.
View Article and Find Full Text PDF